SUNTEST XXL+ FD

The Atlas SUNTEST XXL+ FD is a large state-of-the-art xenon flatbed instrument. It’s completely target designed to meet the testing needs of the FMCG industry, such as drinks, foodstuffs, personal care or household products. XXL+ FD enables consumer goods manufacturers to test their products realistically within very short test times, typically with acceleration factors between 10-30 versus real time. It’s for reliable qualification of stable product formulations, determination of appropriate packaging materials, and supporting product shelflife claims.  

  • 3x 1700 W air-cooled Xenon Lamps
  • 3000 cm2 exposure area
  • Touch screen with colour display to monitor and display test parameters
  • Measurement and Control of Irradiance 300-400 nm / 340nm or 420nm
  • Measurement and Control of Black Standard (BST); (Black Panel optional)
  • Measurement and Control of Chamber Temperature
  • Display of Diagnostic Messages
  • Storage for 10 user-defined test methods
  • Pre-programmed international standard test method
  • Multilingual User Interface
  • Control Panel tiltable for increased operating comfort
  • Thermal insulation of test chamber and periphery preventing water condensation
  • Temperature control selectable either via CHT (15° * - 70 °C; *with chiller) or via CHT and BST (25° * - 100 °C; *with chiller) simultaneously (dual control)
  • Adapted chiller SunCool XXL+ for testing at low temperatures in the range 15-30°C CHT; detachable for standard weathering testing needs
  • Ultrasonic humidification system for controlled humidity during operation without SunCool
  • Specimen spray system with integrated 60 l water reservoir for spray cycles at 25°C ** - 70°C
  • Static, horizontal exposure area for flat or 3D specimen
  • Test chamber with 5 ° slope for optimum spray water drain
  • Access port ca. 3,0 x 2,5 cm for potential use of additional external sensors inside the test chamber
  • Light cassette including non-ageing aluminium reflector, coated quartz filter and slot for additional optical filters
  • Switch-off by radiant exposure, light exposure time or total operating time
  • Advanced Auto Start functions
  • Data output via Memory Card or RS 232
  • Ethernet interface plus XenoTouch Add-ons 1, 2, and 3 (“Remote Control“, “E-Mail Service“, “Online Monitoring“)
  • Sliding sample table for easy specimen loading; exposure area (WxD) 74 x 37 cm
  • Daylight filter
  • Filter set "Solar ID 65"
  • Window Glass Filters
  • XenoTouch Add-ons; Remote Control, E-Mail Service, Online Monitoring
  • SunCal BB 300-400 BST sensor for calibration of irradiance in W/m² in the wavelength range 300-400 nm and for calibration of BST in °C
  • ICH Guideline Q1B
    Guideline for the photostability testing of new drug substances and products
  • ICH Guideline Q5C
    Guideline for Industry - Quality of Biotechnological Products: Stability Testing of Biotechnological/Biological Products
  • ISO 11341
    Paints and Varnishes - Artificial Weathering and Exposure to Artificial Radiation - Exposure to Filtered Xenon-Arc Radiation
  • ISO 4892-2
    Plastics - Methods of exposure to laboratory light sources - Part 2: Xenon-arc sources

Product Video

The SUNTEST Family Line

FMCG Industry - Atlas Can Help You Make the Right Decisions